A-A-53446
Sampling for tests. Sampling for tests including dimensional test, shall be conducted in accordance with MIL-STD-105, with an AQL of 2.5 (percent defective) and inspection level of S-3. The unit of product for test purposes shall be one fountain syringe.
Tests.
Clamp Test. A sample enema unit, with shut-off clamp closed, shall be filled to its nominal capacity with water at 120°F by its suspension device for 60 minutes. The unit shall then be examined for evidence of leakage and suspension failure. Any leakage or suspension shall constitute a defect.
Drop Test. A sample enema unit, with shut-off clamp and filling spout closed, shall be filled to its nominal capacity with warm water at 120°F and dropped from a height of two feet onto a hard regular surface. Any rupture or leak shall constitute a defect.
Contractor certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers’ own drawings, specifications, standards, and quality assurance practices. The Government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract.
Regulatory requirements.
Federal Food, Drug and Cosmetic Act. If the product covered by this document has been determined by the U.S. Food and Drug Administration to be under its jurisdiction, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of Federal Food, Drug and Cosmetic Act, as amended, and regulations promulgated thereunder. In addition, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of all other applicable Federal, State, and local statutes, ordinances, and regulations.
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business