A-A-51355
Electrical conductivity test. The stretcher shall be tested for electrical conductivity according to National Fire Protection Association (NFPA) Standard No. 99, Health Care facilities. The test shall be made from the top of the stretcher body to a metal plate placed successively under each caster. A test failure shall constitute a defect.
Bumper hardness test. The durometer hardness of the stretcher rubber bumper shall be tested using ASTM Method No. D2240-75. A hardness result less than 50 durometer or greater than 70 durometer shall constitute a defect.
Metric products. Products manufactured to metric dimensions will be considered on an equal basis with those manufactured using inch-pound units, providing they fall within specified tolerances using conversion tables contained in the latest revision of Federal Standard 376, and all other requirements of this commercial item description are met.
If a product is manufactured to metric dimensions and those dimensions exceed the tolerances specified in the inch-pound units, a request shall be made to the contracting officer to determine if the product is acceptable.
The contracting officer has the option of accepting or rejecting the product.
Contractor certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers’ own drawings, specifications, standards, and quality assurance practices. The Government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract.
Regulatory requirements .
Federal Food, Drug and Cosmetic Act. If the product covered by this document has been determined by the U.S. Food and Drug Administration to be under its jurisdiction, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of the Federal Food, Drug and Cosmetic Act, as amended, and regulations promulgated thereunder. In addition, the
offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of all other applicable Federal, State, and local statutes, ordinances, and regulations.