A-A-54444
19 July 1991
COMMERCIAL ITEM DESCRIPTION
VENTILATOR, VOLUME, PORTABLE
The General Services Administration has authorized the use of this Commercial Item Description.
This Commercial Item Description covers a self-contained, portable volume ventilator which operates on 11-30 volts DC and also on 110/230 volts AC, 50-400 Hz when used with the separate AC power supply furnished.
Salient Characteristics:
The portable volume ventilator containing time cycling, pressure limiting, constant flow, and microprocessor (CPU) control shall be suitable for use in hospitals, ambulances, fields, and in military aeromedical aircraft. It shall be suitable for providing controlled ventilations to the non-breathing patient and for assisting ventilations to the spontaneously breathing patient.
The ventilator shall be a self-contained apparatus except for a compressed gas source and shall contain at least the following: a control/display panel; internal battery; battery recharge circuit; EMI/RFI suppression circuitry; valve to select flow of gas; alarms; and chart depicting tidal volume by gas flow/inspiratory time. The apparatus with above mentioned items shall be self-contained and have its own separate carrying case with condensed operating instructions marked directly on the case or on a label attached to the case. These instructions are in addition to the specified Operating and Maintenance Manuals. The case shall be semi-rigid plastic or corrosion-resistant metal material with integral carrying case handle, and shall have a hinged, clear door to protect the operator controls of the control/display panel from inadvertent contact and provide splash protection.
DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited.
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business